Author:Makkonen, Joonas
Title:Blood Pressure Measurement Utilizing MEMS Pressure Sensors
Verenpaineen mittaus MEMS-paineantureilla
Publication type:Master's thesis
Publication year:2014
Pages:viii + 51      Language:   eng
Department/School:Sähkötekniikan korkeakoulu
Main subject:Bioniikka   (S3006)
Supervisor:Sepponen, Raimo
Instructor:Meriheinä, Ulf
Electronic version URL: http://urn.fi/URN:NBN:fi:aalto-201403061528
OEVS:
Electronic archive copy is available via Aalto Thesis Database.
Instructions

Reading digital theses in the closed network of the Aalto University Harald Herlin Learning Centre

In the closed network of Learning Centre you can read digital and digitized theses not available in the open network.

The Learning Centre contact details and opening hours: https://learningcentre.aalto.fi/en/harald-herlin-learning-centre/

You can read theses on the Learning Centre customer computers, which are available on all floors.

Logging on to the customer computers

  • Aalto University staff members log on to the customer computer using the Aalto username and password.
  • Other customers log on using a shared username and password.

Opening a thesis

  • On the desktop of the customer computers, you will find an icon titled:

    Aalto Thesis Database

  • Click on the icon to search for and open the thesis you are looking for from Aaltodoc database. You can find the thesis file by clicking the link on the OEV or OEVS field.

Reading the thesis

  • You can either print the thesis or read it on the customer computer screen.
  • You cannot save the thesis file on a flash drive or email it.
  • You cannot copy text or images from the file.
  • You cannot edit the file.

Printing the thesis

  • You can print the thesis for your personal study or research use.
  • Aalto University students and staff members may print black-and-white prints on the PrintingPoint devices when using the computer with personal Aalto username and password. Color printing is possible using the printer u90203-psc3, which is located near the customer service. Color printing is subject to a charge to Aalto University students and staff members.
  • Other customers can use the printer u90203-psc3. All printing is subject to a charge to non-University members.
Location:P1 Ark Aalto  1005   | Archive
Keywords:blood pressure
MEMS
pressure sensor
non-invasive
tonometry
verenpaine
paineanturi
ei-invasiivinen
tonometria
Abstract (eng):The aim of this thesis was to create a device for non-invasive and continuous blood pressure measurement utilizing a MEMS pressure sensor.
The device was designed and produced and its performance was compared with a reference blood pressure monitor in the course of this thesis.
Blood pressure is one of the vital signs of the human body.
Thus its fast and reliable measurement is important to quickly assess the health of a patient.
In addition, long-term high blood pressure, hypertension, is directly related to 6 % of fatalities in the world.
Hypertension has been proven to increase the risk for multiple cardiovascular events, and with an aging population, the total cost of these events will increase.
Thus preventive health care is becoming more and more important.
As a consequence, a method for non-invasive and continuous blood pressure measurement has long been sought after.
In this thesis, a non-invasive, cuff-less blood pressure monitor was produced.
It records pulse waveforms by utilizing tonometry.
From the pulse wave, parameters such as relative amplitudes and timings are calculated.
These parameters are further utilized in determining blood pressure.
The monitor does not require calibration.
In total, pulse wave and reference blood pressures were recorded from 32 subjects.
Re-substitution validation and leave-one-out cross-validation were utilized in assessing the performance of the device.
The errors were 0 +- 6.42 for SBP and +- 5.1 for DBP, and 0 +- 6.42 for SBP and 0 +- 8.11 for DBP respectively.
Abstract (fin):Tämän työn tarkoituksena oli toteuttaa MEMS-paineanturilla laite ei-invasiiviseen ja jatkuvaan verenpaineen mittaamiseen.
Laite suunniteltiin ja rakennettiin ja sen suorituskyky verrattiin referenssiverenpainemittarin vastaavaan.
Verenpaine on yksi kehon elintoimintojen perusmittareista.
Sen nopea ja luotettava mittaus on tärkeää potilaan kunnon määrittämisessä.
Lisäksi korkea verenpaine on suoraan liitettävissä 6 %:iin kuolemantapauksista maailmanlaajuisesti.
Sen on osoitettu lisäävän sydän- ja verensuonitautien riskiä.
Ikääntyvän väestön myötä myös korkean verenpaineen aiheuttamat kustannukset tulevat kasvamaan, mikä lisää ennakoivan terveydenhuollon kysyntää.
Ennakoivassa terveydenhuollossa on tärkeää havaita ajoissa mahdolliset terveydentilaan vaikuttavat tekijät.
Tämän vuoksi tarve ei-invasiiviselle ja jatkuvalle verenpaineen mittaukselle on kasvanut.
Tässä työssä valmistettiin ei-invasiivinen verenpainemittari, joka ei tarvitse mansettia toimiakseen.
Se analysoi pulssiaallon, jonka muodosta lasketaan tiettyjä parametreja, kuten suhteellisia korkeuksia ja ajoituksia.
Näiden laskettujen parametrien avulla estimoidaan verenpaine.
Kehitetty verenpainemittari eroaa vastaavista sillä, ettei se tarvitse erillistä kalibrointia.
Työssä 32 ihmiseltä nauhoitettiin pulssiaalto ja mitattiin verenpaine referenssimittarilla.
Laitteen toiminta validoitiin uudelleenkorvaus- ja ristiinvertailumenetelmien avulla.
Ensimmäisessä malli muodostettiin kaikista koehenkilöistä ja tällä estimoitiin samojen koehenkilöiden verenpaine.
Mittausvirhe oli systoliselle 0 +- 6.42 ja diastoliselle verenpaineelle 0 +- 5.1.
Jälkimmäisessä malli muodostettiin 31 koehenkilöstä ja verrattiin yhteen henkilöön.
Tämä toteutettiin 32 kertaa niin, että jokainen koehenkilö oli vuorollaan vertauksen kohteena.
Tällöin virheeksi saatiin systoliselle 0 +- 6.42 ja diastoliselle verenpaineelle 0 +- 8.11.
ED:2014-03-06
INSSI record number: 48743