Author:Sainio, Sami
Title:Sementtikiinnitteisten lonkkaproteesien polymeeriliukupinnan kuluma ja sen mittaaminen tietokoneavusteisesti
Cement fixed hip prosthesis polymer bearing wear and measurement of it using computerized method
Publication type:Master's thesis
Publication year:2013
Pages:vi + 75      Language:   fin
Department/School:Sähkötekniikan korkeakoulu
Main subject:Bioadaptiivinen tekniikka   (S3006)
Supervisor:Paulasto-Kröckel, Mervi
Instructor:Laurila, Tomi
Electronic version URL: http://urn.fi/URN:NBN:fi:aalto-201304161896
OEVS:
Electronic archive copy is available via Aalto Thesis Database.
Instructions

Reading digital theses in the closed network of the Aalto University Harald Herlin Learning Centre

In the closed network of Learning Centre you can read digital and digitized theses not available in the open network.

The Learning Centre contact details and opening hours: https://learningcentre.aalto.fi/en/harald-herlin-learning-centre/

You can read theses on the Learning Centre customer computers, which are available on all floors.

Logging on to the customer computers

  • Aalto University staff members log on to the customer computer using the Aalto username and password.
  • Other customers log on using a shared username and password.

Opening a thesis

  • On the desktop of the customer computers, you will find an icon titled:

    Aalto Thesis Database

  • Click on the icon to search for and open the thesis you are looking for from Aaltodoc database. You can find the thesis file by clicking the link on the OEV or OEVS field.

Reading the thesis

  • You can either print the thesis or read it on the customer computer screen.
  • You cannot save the thesis file on a flash drive or email it.
  • You cannot copy text or images from the file.
  • You cannot edit the file.

Printing the thesis

  • You can print the thesis for your personal study or research use.
  • Aalto University students and staff members may print black-and-white prints on the PrintingPoint devices when using the computer with personal Aalto username and password. Color printing is possible using the printer u90203-psc3, which is located near the customer service. Color printing is subject to a charge to Aalto University students and staff members.
  • Other customers can use the printer u90203-psc3. All printing is subject to a charge to non-University members.
Location:P1 Ark Aalto  1448   | Archive
Keywords:prosthesis
polymer
bearing couple
wear
measurement accuracy
tekonivel
polymeeri
liukupari
kuluma
mittaustarkkuus
Abstract (eng):Installation rate of prosthetic implants in Finland and worldwide has been and is still increasing.
This is due to population aging and increase in lifespan.
These factors lead to significant costs.
Biological reaction to the wear and wear particles from prosthetic implants depends significantly on the materials used in the bearing couple.
Within the last decade a lot of metal-on-metal bearing couple hip prostheses have been installed to physically active patients.
Wear particles from these bearing couples have been found to be harmful to the patients.
Due to this metal-on-metal hip prostheses have been banned from use.
Metal-on-polymer, ceramic-on-polymer and ceramic-on-ceramic bearing couples also generate wear particles under loading, but the biological effects of these particles are significantly less severe.
In the literature part of this work the reasons leading to total hip replacements, anatomy and biomechanics of the hip as well as loads in the hip joint reported in the literature have been displayed.
By interviewing doctors general guidelines of selecting the materials used in total hip replacements for different patients has been established.
In the experimental part of the work wear measurements using EBRA-method (Einzel Bild Röntgen Analyse) from metal-on-polymer bearing couple are reported.
Scanning electron microscope (SEM) pictures from bearing surfaces are provided as well.
During the work multiple surgeries were followed to understand the clinical side of the implantology.
Wear measurements show that the metal-on-polymer bearing couple suffers from maximum of 0,25 mm linear wear yearly.
Results found are in line with those from the literature.
Based on this study there is no evidence to claim that metal-on-polymer bearing couple prostheses would be insufficient in total hip replacement prostheses.
This work concentrates in the errors found in the wear measurements that have been ignored in many parts in the literature.
Since majority of the wear measurements found from literature are done from AP-radiographs (Anterior-Posterior) without taking in account errors in the measurements, the results in these publications are not as accurate as the publishers report.
In this work reasons for the errors in the measurements are documented and alternative ways of measuring the wear are given.
It can be concluded that EBRA-method is valid tool for measuring wear in metal-on-polymer bearing couples as long as the errors affecting the measurements are taken in account.
It must be noted that functionality of the prosthetic hip depends significantly from the position that it is installed to.
Abstract (fin):Lonkan tekonivelten käyttö on yleistynyt Suomessa ja maailmalla huomattavasti, asennusmäärien ollessa jatkuvasti kasvussa.
Tämä johtuu sekä väestön ikääntymisestä että keskimääräisen eliniän jatkuvasta kasvusta.
Lonkan tekonivelleikkausten yleistyminen johtaa väistämättä merkittäviin taloudellisiin kustannuksiin.
Tekonivelen liukuparin ja sen kulumasta syntyvien kulumapartikkeleiden biologinen vaste vaihtelee merkittävästi materiaaliryhmittäin.
Erityisesti viimeisen vuosikymmenen aikana liikunnallisesti aktiivisille potilaille on asennettu metalli-metalli liukuparillisia lonkan tekoniveliä.
Kyseisten implanttien kulumapartikkelit ovat osoittautuneet elimistölle erittäin haitallisiksi.
Tämä on johtanut metalli-metalli liukuparisten implanttien käyttökieltoon Suomessa ja maailmalla.
Metalli-polymeeri, keraami-polymeeri ja keraami-keraami liukupariset tekonivelet kuluvat myös, mutta niiden kulumapartikkeleiden biologisen vasteen on todettu olevan merkittävästi lievempi.
Työn kirjallisuusosassa on esitetty syyt lonkan tekonivelleikkauksille, esitelty lonkan anatomiaa sekä käyty läpi alan kirjallisuudesta löytyvää lonkan mekaanista kuormitustietoa.
Lisäksi on esitelty yleiset liukuparimateriaalit ja niiden ominaisuudet.
Lääkäreitä on kerätty yleisiä suosituksia eri liukuparien asentamisesta erityyppisille potilaille.
Työn kokeellisessa osassa mitattiin metalli-polymeeriliukuparisten lonkan tekonivelten kulumaa EBRA-menetelmällä (Einzel Bild Röntgen Analyse), sekä kuvattiin liukupintoja pyyhkäisyelektronimikroskoopilla (SEM).
Työn aikana seurattiin myös lukuisia tekonivelleikkauksia, jotta implantologian kliininen puoli tulisi tutuksi.
Mittaustulosten perusteella liukupinnassa havaittiin vuosittain suurimmillaan noin 0,25 mm lineaarinen kuluma.
Tulokset ovat yhtenevät alan kirjallisuudesta löytyvien tulosten kanssa, eikä tämän tutkimuksen perusteella voida esittää syitä siihen, että metalli-polymeeriliukupariset tekonivelet olisivat sopimattomia lonkan tekonivelsovelluksiin.
Tässä työssä tarkastellaan erityisesti kulumamittausjärjestelmien tarkkuutta, joka on usein sivuutettu alan kirjallisuudessa.
Koska lähes kaikki kulumamittaukset on tehty tavallisista AP-röntgenkuvista (Anterior-Posterior) mittaustarkkuus huomioimatta, ovat tulokset huomattavasti epätarkempia kuin alan julkaisuissa esitetään.
Tässä työssä perustellaan syyt mittausten epätarkkuudelle sekä käsitellään vaihtoehtoisia mittausmenetelmiä.
Yhteenvetona voidaan todeta, että työssä on osoitettu EBRA-menetelmän sopivan lonkan tekonivelten kulumamittauksiin, mutta vain jos mittaukseen liittyvät epätarkkuudet otetaan riittävällä tavalla huomioon.
Lopuksi todetaan lonkan tekonivelen toiminnan riippuvan merkittävästi asennosta johon se on asennettu.
ED:2013-05-21