search query: @keyword spectral responsivity / total: 5
reference: 1 / 5
« previous | next »
Author: | Hirvonen, Juha-Matti |
Title: | Spectrally Adjustable Radiance Source |
Spektrisesti säädettävä radianssilähde | |
Publication type: | Licentiate thesis |
Publication year: | 2013 |
Pages: | 40 Language: eng |
Department/School: | Signaalinkäsittelyn ja akustiikan laitos |
Main subject: | Mittaustekniikka (S-108) |
Supervisor: | Ikonen, Erkki |
Instructor: | Kärhä, Petri |
Electronic version URL: | http://urn.fi/URN:NBN:fi:aalto-201308227621 |
Location: | |
Keywords: | LED radiance source spectral responsivity photometry radianssilähde spektrinen herkkyys fotometria |
Abstract (eng): | A spectrally adjustable radiance source based on light emitting diodes (LEDs) has been constructed for spectral responsivity measurements of radiance and luminance meters. A 300-mm integrating sphere source with adjustable output port is illuminated using 30 thermally stabilized narrow-band LEDs covering the visible wavelength range of 380-780 nm. The functionality of the measurement setup is demonstrated by measuring the relative spectral responsivities of a luminance meter and a photometer head with cosine-corrected input optics. |
Abstract (fin): | Työssä esitellään led-pohjainen spektrisesti säädettävä radianssilähde radianssi- ja luminanssimittareiden spektrisen herkkyyden mittaamiseen. Integroivaa palloa, jonka halkaisija on 300 mm, valaistaan 30 kapeakaistaisella ledillä, joiden aallonpituudet kattavat näkyvän valon aallonpituusalueen 380-780 nm ja joiden lämpötila pidetään vakiona. Mittausasetelman toimivuus esitellään mittaamalla suhteellinen spektrinen herkkyys luminanssimittarille ja fotometrimittapäälle, jossa on kosini-korjattu sisääntulo-optiikka. |
ED: | 2013-09-25 |